Introduction: Quantifying inert gas wash-out is crucial to understanding the pathophysiology of decompression sickness. In this study, we developed a portable closed-circuit device for measuring inert gas wash-out and validated its precision and accuracy both with and without human subjects. Methods: We developed an exhalate monitor with sensors for volume, temperature, water vapor and oxygen. Inert gas volume was extrapolated from these inputs using the ideal gas law. The device’s ability to detect volume differences while connected to a breathing machine was analysed by injecting a given gas volume eight times. One hundred and seventy-two coupled before-and-after measurements were then compared with a paired t-test. Drift in measured inert gas volume during unlabored breathing was evaluated in three subjects at rest using multilevel linear regression. A quasi-experimental cross-over study with the same subjects was conducted to evaluate the device’s ability to detect inert gas changes in relation to diving interventions and simulate power. Results: The difference between the injected volume (1,996 ml) and the device’s measured volume (1,986 ml) was -10 ml. The 95% confidence inte