, Taylan Zaman, Abdusselam Celebi, , Bengusu Mirasoglu, · Diving and Hyperbaric Medicine Journal · 2020

Introduction: Noise has physical and psychological effects on humans. Recommended exposure limits are exceeded in many hospital settings; however, information about sound levels in hyperbaric oxygen treatment chambers is lacking. This study measured in-chamber sound levels during treatments in Turkish hyperbaric centres. Methods: Sound levels were measured using a sound level meter (decibel meter). All chambers were multiplace with similar dimensions and shapes. Eight measurements were performed in each of 41 chambers; three during compression, three during decompression, and two at treatment pressure, one during chamber ventilation (flushing) and one without ventilation. At each measurement a sound sample was collected for 25 seconds and A-weighted equivalent (LAeq) and C-weighted peak (LCpeak) levels were obtained. Recorded values were evaluated in relation to sound level limits in regulations. Results: The highest sound level measured in the study was 100.4 dB(A) at treatment pressure while ventilation was underway and the lowest was 40.5 dB(A) at treatment pressure without ventilation. Most centres had sound levels between 70 dB and 85 dB throughout the treatment. Ventilation c