Objective: Elastic impression materials considered favourable decision in fixed and removable prosthesis due to ease of use and accuracy. High price tag is the main drawback. The purpose of this study was to evaluate surface detail reproduction and dimensional stability of newly introduced 3M Preliminary Penta™ VPS in comparison to other elastic impression. Methods: Alginate, alginate replacement and 3M light body were tested for surface detail reproduction and dimensional stability. Ten samples were prepared for each using a stainless-steel die that was created according to ADA specification No. 18 and 19. Surface detail was assessed directly after setting while dimensional stability was evaluated immediately and 24hr after. The data were analysed using a paired sample t-test. Results: All the samples were able to record 75μm line for surface detail. Surface record of 3M imprint™ 4 Preliminary Penta™ VPS showed a similar result to light body. While alginate failed to reproduce 50μm and 20μm lines. Regards to dimensional stability, light body resisted change with respect to time. In contrast, alginate revealed a significant effect (p˂0.05) on dimensional stability for the same stor