Comparison of three dimensional strain volume reconstructions using SOUPR and wobbler based acquisitions: A phantom study

Wenjun Yang, Atul Ingle, Tomy Varghese · Medical Physics · 2016

Purpose:Ultrasound strain imaging is a relatively low cost and portable modality for monitoring percutaneous thermal ablation of liver neoplasms. However, a 3D strain volume reconstruction from existing 2D strain images is necessary to fully delineate the thermal dose distribution. Tissue mimicking (TM) phantom experiments were performed to validate a novel volume reconstruction algorithm referred to as sheaf of ultrasound planes reconstruction (SOUPR), based on a series of 2D rotational imaging planes.Methods:Reconstruction using SOUPR was formulated as an optimization problem with constraints on data consistency with 2D strain images and data smoothness of the volume data. Reconstructed ablation inclusion dimensions, volume, and elastographic signal to noise ratio (SNRe) and contrast to noise ratio (CNRe) were compared with conventional 3D ultrasound strain imaging based on interpolating a series of quasiparallel 2D strain images with a wobbler transducer.Results:Volume estimates of the phantom inclusion were in a similar range for both acquisition approaches. SNRe and CNRe obtained with SOUPR were significantly higher on the order of 250% and 166%, respectively. The mean error o