AbstractBackground and purposeAlthough fundoscopy is a crucial part of the neurological examination, it is challenging, under‐utilized and unreliably performed. The aim was to determine the prevalence of fundus pathology amongst neurology inpatients and the diagnostic accuracy of current fundoscopy practice compared with systematic screening with smartphone fundoscopy (SF) and portable non‐mydriatic fundus photography (NMFP).MethodsThis was a prospective cross‐sectional surveillance and diagnostic accuracy study on adult patients admitted under neurology in an Australian hospital. Inpatients were randomized to initial NMFP (RetinaVue 100, Welch Allyn) or SF (D‐EYE) followed by a crossover to the alternative modality. Images were graded by neurology doctors, using telemedicine consensus neuro‐ophthalmology NMFP grading as the reference standard. Feasibility parameters included ease, comfort and speed.ResultsOf 79 enrolled patients, 14.1% had neurologically relevant pathology (seven, disc pallor; one, hypertensive retinopathy; three, disc swelling). The neurology team performed direct ophthalmoscopy in 6.6% of cases and missed all abnormalities. SF had a sensitivity of 30%–40% compar